dc.description.abstractIntroduction: D-dimer is a crucial test to exclude deep vein thrombosis (DVT). We aimed to evaluate the per- formance of three D-dimer assays: STA-Liatest D-Di Plus (Diagnostica Stago), Tina-quant D-Dimer Gen. 2 (Roche Diagnostics), and INNOVANCE D-Dimer (Siemens Healthineers Diagnostics) in the exclusion of DVT. Methods: Samples (n = 1032) and clinical data were acquired from a prospective outcome study (Rivaroxaban for Scheduled Work-up of Deep Vein Thrombosis – the Ri-Schedule study), which included patients referred to the emergency department with suspected lower-limb DVT. D-dimer was determined with STA-Liatest, and only patients with positive D-dimer values (≥0.5 μg/mL) as stand-alone, were referred for compression ultrasonog- raphy to confirm or exclude DVT. Patients were followed up 90 days after inclusion. Samples were also analyzed with Tina-quant Gen. 2, and INNOVANCE assays. The diagnostic performances of the three assays were investigated. Results: Positive D-dimer (≥0.5 μg/mL) was found in 733 patients (71%) with STA-Liatest, 691 patients (67%) with Tina-quant Gen. 2, and 766 (74%) with INNOVANCE. DVT was confirmed by compression ultrasonography in 196 patients (27%) with positive D-dimer with STA-Liatest. Of those, six (3%) had negative D-dimer (<0.5 μg/ mL) with at least one of the three assays yielding a failure rate of 0.7% with STA-Liatest, 2% with Tina-quant Gen. 2, and 2% with INNOVANCE. The sensitivity and negative predictive value (NPV) for STA-Liatest were 99.0% (95% CI 96.4–99.9) and 99.3% (95% CI 97.4–99.8), for Tina-quant Gen. 2 97.5% (95% CI 94.1–99.2) and 98.5% (95% CI 96.6–99.4), and for INNOVANCE 98.0% (95% CI 94.9–99.0) and 98.5% (95% CI 96.1–99.4) respectively. The efficiency to exclude DVT based on D-dimer as a stand-alone test was 29% (95% CI 26–33) for STA-Liatest, 33% (95% CI 30–37) for Tina-quant Gen. 2, and 26% (95% CI 23–29) for INNOVANCE. Conclusion: STA-Liatest, Tina-quant Gen. 2, and INNOVANCE showed good performances with sensitivity ≥97% and NPV ≥98%. The efficiency to exclude DVT varied and was highest for Tina-quant Gen. 2, whereas the failure rate was lowest for STA-Liatest.en_US
dc.language.isoengen_US
dc.rightsNavngivelse 4.0 Internasjonal*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/deed.no*
dc.titleEvaluation of the diagnostic performance of three D-dimer assays in patients with suspected deep vein thrombosis: STA-Liatest D-Di plus, Tina-quant D-dimer Gen. 2, and INNOVANCE D-dimeren_US
dc.typeJournal articleen_US
dc.typePeer revieweden_US
dc.description.versionpublishedVersionen_US
cristin.ispublishedtrue
cristin.fulltextoriginal
cristin.qualitycode1
dc.identifier.doi10.1016/j.tru.2023.100147
dc.identifier.cristin2182375
dc.source.journalThrombosis Updateen_US
dc.source.volume13en_US
dc.source.pagenumber5en_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record

Navngivelse 4.0 Internasjonal
Except where otherwise noted, this item's license is described as Navngivelse 4.0 Internasjonal