Evaluation of the diagnostic performance of three D-dimer assays in patients with suspected deep vein thrombosis: STA-Liatest D-Di plus, Tina-quant D-dimer Gen. 2, and INNOVANCE D-dimer
Tønne, Brita; Pedersen, Marit Holmefjord; Frønæs, Synne Grønvold; Jørgensen, Camilla Tøvik; Amundsen, Erik Koldberg; Mæhlum, Julie Berge; Berg, Thea; Mathisen, Åse-Berit; Ghanima, Waleed; Garabet, Lamya Samir Noori
Journal article, Peer reviewed
Published version
Date
2023Metadata
Show full item recordCollections
Abstract
Introduction: D-dimer is a crucial test to exclude deep vein thrombosis (DVT). We aimed to evaluate the per- formance of three D-dimer assays: STA-Liatest D-Di Plus (Diagnostica Stago), Tina-quant D-Dimer Gen. 2 (Roche Diagnostics), and INNOVANCE D-Dimer (Siemens Healthineers Diagnostics) in the exclusion of DVT.
Methods: Samples (n = 1032) and clinical data were acquired from a prospective outcome study (Rivaroxaban for Scheduled Work-up of Deep Vein Thrombosis – the Ri-Schedule study), which included patients referred to the emergency department with suspected lower-limb DVT. D-dimer was determined with STA-Liatest, and only patients with positive D-dimer values (≥0.5 μg/mL) as stand-alone, were referred for compression ultrasonog- raphy to confirm or exclude DVT. Patients were followed up 90 days after inclusion. Samples were also analyzed with Tina-quant Gen. 2, and INNOVANCE assays. The diagnostic performances of the three assays were investigated.
Results: Positive D-dimer (≥0.5 μg/mL) was found in 733 patients (71%) with STA-Liatest, 691 patients (67%) with Tina-quant Gen. 2, and 766 (74%) with INNOVANCE. DVT was confirmed by compression ultrasonography in 196 patients (27%) with positive D-dimer with STA-Liatest. Of those, six (3%) had negative D-dimer (<0.5 μg/ mL) with at least one of the three assays yielding a failure rate of 0.7% with STA-Liatest, 2% with Tina-quant Gen. 2, and 2% with INNOVANCE. The sensitivity and negative predictive value (NPV) for STA-Liatest were 99.0% (95% CI 96.4–99.9) and 99.3% (95% CI 97.4–99.8), for Tina-quant Gen. 2 97.5% (95% CI 94.1–99.2) and 98.5% (95% CI 96.6–99.4), and for INNOVANCE 98.0% (95% CI 94.9–99.0) and 98.5% (95% CI 96.1–99.4) respectively. The efficiency to exclude DVT based on D-dimer as a stand-alone test was 29% (95% CI 26–33) for STA-Liatest, 33% (95% CI 30–37) for Tina-quant Gen. 2, and 26% (95% CI 23–29) for INNOVANCE. Conclusion: STA-Liatest, Tina-quant Gen. 2, and INNOVANCE showed good performances with sensitivity ≥97% and NPV ≥98%. The efficiency to exclude DVT varied and was highest for Tina-quant Gen. 2, whereas the failure rate was lowest for STA-Liatest.