Lesion detection performance : comparative analysis of low-dose CT data of the chest on two hybrid imaging systems
Jessop, Maryam; Thompson, John D.; Sil, Joanne; Sanderud, Audun; Jorge, Jose; De groot, Martin; Lança, Luis; Hogg, Peter
Journal article, Peer reviewed
This research was originally published in j n m. jessop, m., thompson, j. d., sil, j., sanderud, a., jorge, j., de groot, m. ... hogg, p. ( under utgivelse). lesion detection performance : comparative analysis of low-dose c t data of the chest on two hybrid imaging systems. journal of nuclear medicine technology © by the society of nuclear medicine and molecular imaging, inc.
Permanent lenke
https://hdl.handle.net/10642/2448Utgivelsesdato
2015Metadata
Vis full innførselSamlinger
Originalversjon
Jessop, M., Thompson, J.D., Sil, J., Sanderud, A., Jorge, J., De groot, M. ... Hogg, P. (Under utgivelse). Lesion detection performance : comparative analysis of low-dose CT data of the chest on two hybrid imaging systems. Journal of Nuclear Medicine Technology http://dx.doi.org/10.2967/jnmt.114.147447Sammendrag
Purpose: Incidental findings, discovered in low-dose CT images obtained during hybrid imaging, are an increasing
phenomenon with advancing CT technology. Understanding their diagnostic value along with technical
limitations is therefore important when reporting images and recommending follow-up, which may result in
additional radiation dose from further diagnostic imaging and an increase in patient anxiety. This study assesses
lesion detection in CT images obtained during attenuation correction (AC) acquisitions on two SPECT/CT systems.
Methods: An anthropomorphic chest phantom, containing simulated lesions of varying size and density, was
imaged on a GE Infinia Hawkeye 4 and a Siemens Symbia T6 with low-dose CT settings used during AC
acquisitions in myocardial perfusion imaging. Twenty-two readers completed a lesion detection task, assessing
46 images (15 normal, 31 abnormal containing 41 lesions) from each SPECT/CT system. Data was evaluated using
a jackknife alternative free-response receiver operating characteristic (JAFROC) analysis.
Results: JAFROC analysis showed a significant difference (p<0.0001) in lesion detection with figures of merit
0.599 (95% CI 0.568, 0.631) and 0.810 (95% CI 0.781, 0.839) for GE Infinia Hawkeye 4 and Siemens Symbia T6
respectively. Lesion detection on the Infinia Hawkeye 4 was generally limited to larger, higher density lesions.
The Siemens Symbia T6 images allowed improved detection rates with mid-sized lesions and some lower density
lesions. However, readers struggled to detect small (5mm) lesions on both image sets, irrespective of density.
Conclusions: Lesion detection is more reliable in low-dose CT images from the Symbia T6 than those from the
Infinia Hawkeye4. This phantom based study gives an indication of potential lesion detection in the clinical
context as shown by two commonly used SPECT/CT systems, which may assist the clinician in determining if
further diagnostic imaging is justified